| Model Number |
MAJ-209 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject device referenced in this report has not yet been returned to olympus for evaluation.Therefore the exact cause of the reported event could not be conclusively determined at this time.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
|
| |
|
Event or Problem Description
|
|
During an endoscopy, the subject device was used.The intended procedure was completed with the subject device.There was no patient injury reported.The suction connector of the device broke off when the user removed the device after the procedure.This is the report regarding the breakage of the suction connector.
|
| |
|
Additional Manufacturer Narrative
|
|
The subject device was not returned.Therefore, the root cause was assumed based on the past similar cases.It was assumed that the breakage occurred at the proximal end of the suction connector.The manufacturing process and a molding supplier for the subject product were changed.After these changes, the complaints of the suction valve breakage has been raising.Olympus medical systems corp.(omsc) assumes that the cause of the damage to the product is that an unexpected large load was applied to the suction connector.Both "product before changes" and "product after changes" meet the product specification for durability.However, when an unexpected large load was applied to the suction connector, omsc confirmed that the product before changes does not break, but the product after changes may break.Omsc has started to re-produce the product which before changes.The instruction manual of the product has already warned as follows; attaching the suction valve to the endoscope.Press down on the suction valve's top surface with your both thumbs until it "clicks" into place.Inspection of operation.Place the distal end of the insertion tube in sterile water and depress the suction valve.Confirm that water is continuously aspirated into the suction bottle of the suction pump.Removing the suction valve from the endoscope rotate the suction connector clockwise with your thumb.
|
| |
|
Search Alerts/Recalls
|