Additional evaluation code: 19/cause traced to user.In telephone contact, it was informed that the dentist did not have information about lot number of the product.Considering the surgical methodology, it was seen that the dentist has used sequence of drills different than the one recommended for the implant installation.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form and visual conference of the product returned by the dentist.
|
|
(b)(4).The dentist reported that almost 2 months after the dental implant was installed in intraoral region 8#, its non-osseointegration was observed.Moreover, 45n.Cm of primary stability was achieved, the patient presented bone type iii, immediate implant was performed and immediate load procedure was done.
|