• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH VIPER PRIME INSERTER CARRIER; ORTHOPAEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH VIPER PRIME INSERTER CARRIER; ORTHOPAEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 286750033
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 11/29/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #: (b)(4).Complainant part is expected to be returned for manufacturer review/investigation.(b)(4).A review of the device history record has been requested.The investigation could not be completed; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that a scrub nurse had difficulty loading implant.Surgeon was about to implant but noticed the cage was not straight.Interface with inserter was not straight and in alignment with inserter.We tried another concorde bullet inserter, but the cage was still not straight.Had to open another implant which loaded correctly.The surgeon was getting s1 nerve root irritation with neuromonitoring so implanted and removed second implant.Had to get ct scan and navigate remaining screws.It was reported that the implant was faulty.Implant was decontaminated.The viper prime screwdriver was not able to release screw and stylet was bent and had to be cut.Surgery was delayed due to the reported event 10 minutes, no fragments were generated.Concomitant device reported: unknown screws (part# unknown, lot# unknown, quantity unknown).This complaint involves two (2) devices.This report is 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: a review of the receiving inspection (ri) for the viper prime inserter carrier was conducted identifying that lot number: mf4283703 was released in a single batch.Batch1: lot qty of (b)(4) units were released on 03-nov-2017 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Visual inspection: a viper prime inserter assembly was returned to us customer quality (cq), all four of its main components were returned assembled to one another.The assembly consisted of: viper prime inserter carrier (part#: 286750033 / lot#: mf4283703), viper prime insert drive tube (part#: 286750034 / lot#: ), viper prime inserter handle (part#: 286750032, lot#: mf4292702) and viper prime inserter shaft (part#: 286750031, lot#: mf4302702).The prime stylet depth adjustor and viper prime stylet were also received assembled into the inserter assembly.There was no damage evident on the exterior of the assembly.The inserter carrier had no damage to note.Functional test: when rotating the handle, the stylet was successfully withdrawn or extended.The inserter assembly was successfully disassembled by loosening the set screw, and then separating the individual components.Each component was successfully disassembled from one another thus the overall complaint was not confirmed.There was no evidence of a device disassembly issue.Can the complaint be replicated with the returned device(s)? no; device disassembled successfully.Dimensional inspection: dimensional inspection was not performed as the reported condition could not be confirmed and no damage could be identified that would contribute to the complaint condition.Document/specification review: drawing(s) reviewed: current & manufactured revisions.Conclusion: the overall complaint was not confirmed for the received viper prime inserter carrier as the device was successfully disassembled.The handle successfully withdrew/extended the stylet when rotated.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.H6: code 3191 used to capture nerve injury, device failure and surgical intervention.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6: a viper prime inserter assembly was returned to us cq, all four of its main components were returned assembled to one another.The assembly consisted of: viper prime inserter carrier (part # 286750033 / lot # mf4283703), viper prime insert drive tube (part # 286750034 / lot # ), viper prime inserter handle (part # 286750032, lot # mf4292702) and viper prime inserter shaft (part # 286750031, lot # mf4302702).The prime stylet depth adjustor and viper prime stylet were also received assembled into the inserter assembly.There was no damage evident on the exterior of the assembly.The inserter carrier had no damage to note.When rotating the handle, the stylet was successfully withdrawn or extended.The inserter assembly was successfully disassembled by loosening the set screw, and then separating the individual components.Each component was successfully disassembled from one another thus the overall complaint was not confirmed.There was no evidence of a device disassembly issue.Dimensional inspection was not performed as the reported condition could not be confirmed and no damage could be identified that would contribute to the complaint condition.The overall complaint was not confirmed for the received viper prime inserter carrier as the device was successfully disassembled.The handle successfully withdrew/extended the stylet when rotated.There was no indication that a design or manufacturing issue contributed to the complaint.No design issues were observed during the document/specification review.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device history lot a review of the receiving inspection (ri) for the viper prime inserter carrier was conducted identifying that lot number mf4283703 was released in a single batch.¿ batch1: lot units were released on 03-nov-2017 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER PRIME INSERTER CARRIER
Common Device Name
ORTHOPAEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key9727624
Report Number1526439-2020-00525
Device Sequence Number19565143
Product Code OLO
UDI-Device Identifier10705034507606
UDI-Public(01)10705034507606
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K170937
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number286750033
Device Catalogue Number286750033
Device Lot NumberMF4283703
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2020
Initial Date Received by Manufacturer 02/11/2020
Supplement Date Received by Manufacturer02/04/2020
03/31/2020
Initial Report FDA Received Date02/19/2020
Supplement Report FDA Received Date02/24/2020
03/31/2020
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN SCREWS
Outcome Attributed to Adverse Event Required Intervention;
-
-