Model Number 71365607
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Code Injury (2348)
Date of Event 01/22/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a thr procedure, the size 7 broach went deeper down the femur than the size 6.No sign of fracture.The surgeon thinks the broach may be defective.No delays reported.A backup device was available.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The medical investigation concluded that no relevant supporting clinical information has been provided to assist with a clinical investigation, and the patient's current condition is unknown.Therefore based on insufficient information, a thorough clinical assessment cannot be performed at this time.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Some potential probable causes of this event could include improper device size used or a procedural error.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY STD BROACH SIZE 7
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9728785
Report Number1020279-2020-00636
Device Sequence Number16689204
Product Code JDG
UDI-Device Identifier03596010560025
UDI-Public03596010560025
Combination Product (Y/N)N
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number71365607
Device Catalogue Number71365607
Device Lot Number16AM11846
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 01/22/2020
Supplement Date Received by Manufacturer03/26/2020
Initial Report FDA Received Date02/19/2020
Supplement Report FDA Received Date04/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-