• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 977C165
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 02/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: product id: 977c165, serial#: (b)(4), product type: lead.A supplemental report will be submitted when analysis results are available.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a health care provider (hcp) via a manufacturer representative regarding a patient who was implanted with a neurostimulator.It was reported that the patient was in for a full system replacement to update to mri scan models.During lead manipulation of the new leads in epidural space, the electrode caught tissue forceps and lifted off the leads.It was reported that a contributing factor was scar tissue in epidural space from explanted leads.Different lead models were opened and were used successfully and fit in the space with ease.No further complications were reported.
 
Additional Manufacturer Narrative
The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.Analysis identified that the #1 electrode(s) was/were pulled out/off of the distal end of the lead.No shorts between circuits.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECIFY SURESCAN
Common Device Name
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key9729786
Report Number2649622-2020-03654
Device Sequence Number19453502
Product Code LGW
UDI-Device Identifier00643169464599
UDI-Public00643169464599
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/19/2023
Device Model Number977C165
Device Catalogue Number977C165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/12/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/04/2020
Supplement Date Received by Manufacturer04/28/2020
04/30/2020
10/06/2020
Initial Report FDA Received Date02/19/2020
Supplement Report FDA Received Date04/30/2020
05/05/2020
10/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age65 YR
-
-