| Model Number |
977C165 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: product id: 977c165, serial#: (b)(4), product type: lead.A supplemental report will be submitted when analysis results are available.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a health care provider (hcp) via a manufacturer representative regarding a patient who was implanted with a neurostimulator.It was reported that the patient was in for a full system replacement to update to mri scan models.During lead manipulation of the new leads in epidural space, the electrode caught tissue forceps and lifted off the leads.It was reported that a contributing factor was scar tissue in epidural space from explanted leads.Different lead models were opened and were used successfully and fit in the space with ease.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned device was subjected to a series of standard tests that include but is not limited to visual inspection and electrical testing.Analysis identified that the #1 electrode(s) was/were pulled out/off of the distal end of the lead.No shorts between circuits.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|