• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US HEALIX AWL -NS; BONE AWL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US HEALIX AWL -NS; BONE AWL Back to Search Results
Model Number 222223
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 02/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).Device was used for treatment, not diagnosis.Udi: (b)(4).
 
Event or Problem Description
It was reported by the sales rep via complaint submission tool that when performing a shoulder rotator cuff repair when the healix awl ns was being used it was noticed that the laser line was completely dull and was not able to properly locate the depth to mallet.A second 4.75 healix advance kntlss pk that the facility owned was intended to be used however the top of the awl was slightly bent.The procedure was completed using another device.No patient consequence and no surgical delay was reported.No additional information was provided.An email received from the affiliate clarified the second product code 222223, lot 1408002, healix awl.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary according to the information provided, it was reported that during a shoulder rotator cuff repair when the healix awl -ns was being used it was noticed that the laser line was completely dull and was not able to properly locate the depth to mallet.A second 4.75 healix advance kntlss pk that the facility owned was intended to be used however the top of the awl was slightly bent.The complaint device is not being returned, therefore unavailable for a physical evaluation.Since the complaint device was not returned, we cannot determine a root cause for the reported failure.If the device is received in the future, we will reopen the complaint and perform the investigation as appropriate.  a manufacturing record evaluation was performed for the finished device lot/serial number (b)(6), and no non-conformances were identified at this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Device history lot a manufacturing record evaluation was performed for the finished device lot/serial number (b)(6), and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEALIX AWL -NS
Common Device Name
BONE AWL
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key9732460
Report Number1221934-2020-00626
Device Sequence Number3098406
Product Code LXH
UDI-Device Identifier10886705007707
UDI-Public10886705007707
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number222223
Device Catalogue Number222223
Device Lot Number1408002
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 02/07/2020
Supplement Date Received by Manufacturer04/06/2020
Initial Report FDA Received Date02/20/2020
Supplement Report FDA Received Date04/07/2020
Patient Sequence Number1
-
-