Model Number 303511
Medical Device Problem Codes Failure to Osseointegrate (1863); Osseointegration Problem (3003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 11/06/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event has been determined to be a post-placement/pre-load failure.This implant loss suggests that the source of the problem was likely to stem from mobility and no osseointegration.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, overall health or behavior.Proper intended use of the implant is described in its instructions for use.
 
Event or Problem Description
Inadequate bone quality/bone quantity and mobility were reported.Time of loss in relation to the implantation was before prosthetic restoration.Bone quality at the time of implant failure type iii.Primary stability and osseointegration was not achieved.Augmentation procedure (allograft) was performed at the time of surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRICAM INTERNAL IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ACE SURGICAL SUPPLY COMPANY
1034 pearl street
brockton MA 02301
MDR Report Key9733036
Report Number1287163-2020-00699
Device Sequence Number353376
Product Code DZE
UDI-Device Identifier00614950005185
UDI-Public00614950005185
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K103790
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number303511
Device Catalogue Number303511
Device Lot Number18050074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/28/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/28/2020
Initial Report FDA Received Date02/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age55 YR
-
-