• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. DISCOVERY ELBOW; DISC ULNA 3X115MM RT W BRNG C

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. DISCOVERY ELBOW; DISC ULNA 3X115MM RT W BRNG C Back to Search Results
Catalog Number 114817
Device Problem Loose or Intermittent Connection (1371)
Patient Problem Failure of Implant (1924)
Event Date 01/22/2020
Event Type  Injury  
Manufacturer Narrative
Additional reporting on this event will be provided as a supplemental report to these documents as soon as it becomes available.
 
Event Description
Revision surgery - ulna component became loose.
 
Manufacturer Narrative
The reason for this revision surgery was reported as loosening.The actual length of in-vivo for the items listed is unknown as the original surgery date was not provided or could be established.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The devices were disposed of at hospital and not made available to djo surgical for examination.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the products that may have contributed to the reported event.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to loosening.There were no findings during this evaluation that indicate the reported devices were defective.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISCOVERY ELBOW
Type of Device
DISC ULNA 3X115MM RT W BRNG C
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445,
MDR Report Key9733269
MDR Text Key180346008
Report Number1644408-2020-00160
Device Sequence Number1
Product Code JDC
Combination Product (y/n)N
PMA/PMN Number
K013042
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 03/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date11/03/2022
Device Catalogue Number114817
Device Lot Number530920
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/22/2020
Initial Date FDA Received02/20/2020
Supplement Dates Manufacturer Received02/21/2020
Supplement Dates FDA Received03/19/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
114700 LOT 216410
Patient Outcome(s) Required Intervention;
Patient Age80 YR
-
-