| Catalog Number |
0377031610 |
| Medical Device Problem Codes |
Break (1069); Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
01/16/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the arthroscopic cannula broke during procedure & pieces potentially remained in joint.Please note that the procedure was completed successfully.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: torn arthroscopic cannula probable root cause: poor autoclave reliability use error the reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.
|
| |
|
Event or Problem Description
|
|
It was reported that the tip of the arthroscopic cannula broke during procedure & pieces potentially remained in joint.Please note that the procedure was completed successfully.
|
| |
|