Model Number 109.669
Medical Device Problem Code Failure to Osseointegrate (1863)
Health Effect - Clinical Code Tissue Damage (2104)
Date of Event 12/18/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Considering the surgical methodology, it was seen that the dentist has used sequence of drills different than the one recommended for the implant installation.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form, visual conference of the product returned by the dentist and the evaluation of relevant production records.
 
Event or Problem Description
(b)(4).The dentist reported that 20 days after the dental implant was installed in intraoral regio ada 22, its non-osseointegration was observed.Moreover, the patient presented bone type iii, bone graft was placed, immediate implant was performed and immediate load procedure was done.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CM TITAMAX EX IMPLANT 3.75X15 MM
Common Device Name
ENDOSSEOUS DENTAL IMPLANT
Manufacturer (Section D)
NEODENT - JJGC S.A.
av. juscelino kubitschek de
oliveira, 3291 cic
curitiba PR 81270 -200
BR  81270-200
MDR Report Key9734058
Report Number3008261720-2020-00924
Device Sequence Number366971
Product Code DZE
UDI-Device Identifier07898237569161
UDI-Public07898237569161
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K101945
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 02/20/2020
Report Date (Section F) 01/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/14/2022
Device Model Number109.669
Device Catalogue Number109.669
Device Lot Number800331345
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/27/2020
Was the Report Sent to FDA? No
Distributor Facility Aware Date02/20/2020
Device Age24 MO
Event Location Other
Date Report to Manufacturer01/27/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/27/2020
Initial Report FDA Received Date02/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-