• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION QUANTUM MAVERICK; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 7015
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 12/07/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device was evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed multiple kinks along the hypotube.The hypotube is separated 73.1cm from the hub.Microscopic examination revealed no additional damages.The balloon is still tightly folded and there is blood present in the guidewire lumen.Inspection of the remainder of the device presented no other damage or irregularities.
 
Event or Problem Description
Reportable based on device analysis completed on 31jan2020.It was reported that shaft kink occurred.The 95% stenosed target lesion was located in the left anterior descending artery.A 4.0mm x 15mm quantum maverick balloon catheter was advanced for dilatation; however, during procedure, the balloon delivery shaft was kinked.The device was still intact when it was removed from the patient, and was also still intact when packed for return to bsc.The procedure was completed with another of same device.No patient complications nor injuries were reported.However, returned device analysis revealed a hypotube separation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUANTUM MAVERICK
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9734384
Report Number2134265-2020-01862
Device Sequence Number11362992
Product Code LOX
UDI-Device Identifier08714729392828
UDI-Public08714729392828
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
P860019
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 02/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/07/2022
Device Model Number7015
Device Catalogue Number7015
Device Lot Number0023757895
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/08/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/31/2020
Initial Report FDA Received Date02/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age86 YR
Patient Weight60
-
-