• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AZIYO BIOLOGICS, INC. CANGAROO ENVELOPE; MESH, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AZIYO BIOLOGICS, INC. CANGAROO ENVELOPE; MESH, SURGICAL Back to Search Results
Model Number CMCV-009-LRG
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 07/15/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturing review of the device history record for the reported lot shows that all units were quality released on 24-apr-2019 having met all internal qc acceptance requirements.All sterilization records, and bioburden testing indicate successful sterilization process, and passing lal and product sterility results allowed the lot to be released having met all criteria for manufacturing release.There were no non-conformances associated with the manufacturing lot during production, sterilization and final packaging.In lieu of a request for the oem supplier dhr review, it is noted that aziyo processes the non-sterile envelope materials by cutting, suturing, packaging, and sterilizing.It can be noted that per the instructions for use [ifu] (part number-20663 provided with device) for the cangaroo envelope currently lists infection as a potential complication associated with the procedure and device.It can also be noted that the cangaroo envelope remains implanted and only the right atrial lead was explanted to facilitate a system revision.Although the exact cause of the reported event cannot be conclusively determined, infection is a known complication with the cangaroo envelope.
 
Event or Problem Description
It was reported that the right atrial (ra) lead was part of a system revision due to infection.There were no additional adverse patient effects reported.The ra lead was explanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANGAROO ENVELOPE
Common Device Name
MESH, SURGICAL
Manufacturer (Section D)
AZIYO BIOLOGICS, INC.
1100 old ellis road
ste 1200
roswell, ga
MDR Report Key9734531
Report Number3005619880-2020-00024
Device Sequence Number2352524
Product Code FTM
UDI-Device Identifier00859389005065
UDI-Public00859389005065
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K182255
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 02/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/14/2021
Device Model NumberCMCV-009-LRG
Device Lot NumberM19D1172077
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 01/27/2020
Initial Report FDA Received Date02/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age75 YR
-
-