MEDTRONIC CRYOCATH LP CRYOCONSOLE; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Model Number 106A3
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/23/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product event summary: data files were returned and analyzed.No patient file was created on the reported date of the event.A power up error message for the reported date of the event did not show any error.The reported start up and power issues are not likely to cause or contribute to a death or serious injury; however, the risk of the patient being under general anesthesia without full therapeutic effect is the adverse event being reported.In conclusion, the decision to abort the procedure without use of alternate therapy was based upon the medical judgment of the physician.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that prior to a cryo ablation procedure, the console would not power on.The case was aborted while the patient was under general anesthesia.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Correction: h6 product event summary: the power supply 11002-098 from cryo console with serial number n-6003 was returned for analysis.Visual inspection showed that the power supply was intact with no apparent issues.The part was assembled to a universal test console, but the console could not be powered up.A ¿no input detection¿ message showed on the screen.Electrical measurement of the power supply did not show any output voltage.In conclusion, the power supply failed the returned product inspection due to the depleted voltage output, however, if the malfunction were to re-occur, it is not likely to cause or contribute to a death or serious injury.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRYOCONSOLE
Common Device Name
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key9734892
Report Number3002648230-2020-00106
Device Sequence Number18924038
Product Code LPB
UDI-Device Identifier00613994686077
UDI-Public00613994686077
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
P020045
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/10/2015
Device Model Number106A3
Device Catalogue Number106A3
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/06/2020
Initial Date Received by Manufacturer 01/23/2020
Supplement Date Received by Manufacturer03/12/2020
Initial Report FDA Received Date02/20/2020
Supplement Report FDA Received Date03/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-