Model Number 502-11-52E
Medical Device Problem Codes Failure to Osseointegrate (1863); Osseointegration Problem (3003)
Health Effect - Clinical Codes Injury (2348); Inadequate Osseointegration (2646)
Date of Event 01/24/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
It was reported that the patient's hip was revised.As reported by rep: "loose acetabular shell".The shell, head and liner were revised.Update per response: "there are no allegations from surgeon against the 2 implants questioned" (revised liner and revised femoral head).
 
Additional Manufacturer Narrative
Reported event: an event regarding loosening involving trident shell was reported.The event was confirmed via medical review.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: a review of the provided medical records by a clinical consultant stated the following comment: provided revision op report and consultation note confirm event, need additional information; primary operative report, clinical and past medical history, additional serial dated x-rays and examination of explanted components.Device history review: could not be performed as lot code information was not provided.Complaint history review: could not be performed as lot code information was not provided.Conclusion: a review of the provided medical records by a clinical consultant stated the following comment: provided revision op report and consultation note confirm event, need additional information; primary operative report, clinical and past medical history, additional serial dated x-rays and examination of explanted components.No further investigation for this event is possible at this time as no devices and/ or insufficient information was received by stryker orthopaedics.If devices and / or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported that the patient's hip was revised.As reported by rep: "loose acetabular shell".The shell, head and liner were revised.Update per response: "there are no allegations from surgeon against the 2 implants questioned" (revised liner and revised femoral head).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENT HEMISPHERICAL CLUSTER HOLE SHELL
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9739323
Report Number0002249697-2020-00370
Device Sequence Number2483125
Product Code LPH
UDI-Device Identifier04546540432032
UDI-Public04546540432032
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K143085
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number502-11-52E
Device Catalogue Number502-11-52E
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 01/24/2020
Supplement Date Received by Manufacturer03/05/2020
Initial Report FDA Received Date02/21/2020
Supplement Report FDA Received Date04/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age63 YR
Patient Weight68
-
-