U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION SMARTSITE VIALSHIELD CLOSED VIAL ACCESS DEVICE; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION SMARTSITE VIALSHIELD CLOSED VIAL ACCESS DEVICE; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number MV0520
Medical Device Problem Codes Fluid/Blood Leak (1250); Material Separation (1562)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Concomitant medical products: paclitaxel injection vial, lot: 9az5231, exp: 06/2022.The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
 
Event or Problem Description
It was reported that the closed vial access device (cvad), attached to a paclitaxel 50ml vial, separated while the pharmacy technician was unscrewing the texium syringe from the cvad.The leak did not cause harm to the pharmacy technician.There was no patient involvement as the event occurred while the medication was being prepared in the pharmacy department.
 
Additional Manufacturer Narrative
The customer¿s report that the closed vial access device (cvad) separated while the pharmacy technician was unscrewing the texium syringe was confirmed by visual inspection.The received products were visually inspected for obvious damage such as incomplete bonding engagements, cracks, fractures or other anomalies.The smartsite component which should be bonded to the cvad had separated from the cvad and was still connected to the texium syringe.There were no physical signs of breakage or stress on the cvad or smartsite.However, the threads of the smartsite¿s female luer adapter were turned upwards at the edges indicating over-tightening.Investigation found marginal solvent between the male/female threads of the mating parts.Functional testing was not performed due to the obvious damage on the received vad.The root cause for the separation could not be definitively determined.Yukon medical device history record review: a device history record review was completed for mv0520 lot 9242.There were no nmrs, capas/deviations generated or reported anomalies during the production of this lot.The date of manufacture is 10/2019.
 
Event or Problem Description
It was reported that the closed vial access device (cvad), attached to a paclitaxel 50ml vial, separated while the pharmacy technician was unscrewing the texium syringe from the cvad.The leak did not cause harm to the pharmacy technician.There was no patient involvement as the event occurred while the medication was being prepared in the pharmacy department.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SMARTSITE VIALSHIELD CLOSED VIAL ACCESS DEVICE
Common Device Name
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
MDR Report Key9740150
Report Number9616066-2020-00570
Device Sequence Number3098518
Product Code FPA
UDI-Device Identifier10885403228117
UDI-Public10885403228117
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
K132863
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMV0520
Device Catalogue NumberMV0520
Device Lot Number9242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2020
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date02/21/2020
Supplement Report FDA Received Date08/03/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2000E,MY8020, THERAPY DATE UNKNOWN
-
-