Model Number 9600TFX23J
Device Problems Obstruction of Flow (2423); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Occlusion (1984); Patient Problem/Medical Problem (2688)
Event Date 01/29/2020
Event Type  Injury  
Manufacturer Narrative
Udi number: (b)(4).Per the instructions for use (ifu), coronary flow obstruction is a potential adverse event associated with the tavr procedure.The ifu cautions that the safety and effectiveness have not been established for patients with bulky calcified aortic valve leaflets in close proximity to coronary ostia.Coronary occlusion can result in myocardial ischemia or infarction due to obstruction of the coronary blood flow and may require intervention (e.G.Pci).There are multiple patient factors that could contribute to coronary occlusion by the prosthetic valve or native valve leaflets, including a minimal distance between the native annulus and the coronary ostia, bulky calcification, long native leaflets, and obliterated coronary sinuses.Procedural factors such as torn native leaflet during bav, plaque shift, deployment of the bioprosthetic heart valve too aortic and significant valve over sizing could also contribute to this complication.The edwards thv training manuals advise the operator on pre-procedure assessment of the aortic valve, root, and coronary anatomy.Physicians are extensively trained by edwards before they are qualified to use the transcatheter heart valve (thv).Training includes patient screening, device preparation, approach, deployment, imaging, procedure-specific training manuals and proctored procedures.The physician is instructed to evaluate this risk early in the patient screening process in all patients the following factors should be considered: degree of calcification on leaflets, annulus to coronary ostia distance, length of the valve leaflet, width of the valsalva sinuses, movement of the leaflets during bav, patency of coronaries during bav, and expanded height of the intended thv.The training manuals also provide the following tips for detecting risk for left main occlusion: (1) aortogram or tee prior to thv implantation to reveal bulky calcified leaflets; (2) during pre-dilatation, note bulky calcification on valve moving towards ostium on left main; and (3) consider aortogram during valvuloplasty to assess coronary flow.In this case, there was no allegation or indication a device malfunction contributed to this adverse event.The patient had a low lmt ostium height of 7.8 mm.Due to risk of coronary occlusion, a coronary stent was preventively deployed.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required.
 
Event Description
As reported by our affiliates in (b)(6), after deployment of a 23mm sapien 3 valve in the aortic position via transfemoral approach, a native leaflet was close to the ostium of the left main trunk (lmt).Distortion of the ostium was confirmed by intravascular ultrasound (ivus).Therefore, a coronary stent was preventively deployed.The patient had no wave change on electrocardiogram (ecg) and no ischemic symptom were observed post valve implant.The lmt ostium height was 7.8 mm.The sinus of valsalva (sov) was narrow at 25-27 mm.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE (THV)
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
MDR Report Key9741246
MDR Text Key190131629
Report Number2015691-2020-10636
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 01/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/23/2021
Device Model Number9600TFX23J
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 01/29/2020
Initial Date FDA Received02/21/2020
Supplement Dates Manufacturer Received07/23/2020
Supplement Dates FDA Received11/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-