Model Number L20416
Medical Device Problem Codes Device-Device Incompatibility (2919); Scratched Material (3020)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.(b)(4).
 
Event or Problem Description
It was reported that the surgeon performed a direct anterior approach total hip replacement on the left side with a pinnacle cup and a corail stem.The surgeon broached up to a size 16 corail broach.When he wanted to calcar plane, he realized the calcar planer did not seat fully over the nub on the broach.After opening another tray with another calcar planer, it was determined that something must be wrong with the size 16 corail broach.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BROACH CORAIL AMT 16
Common Device Name
HIP INSTRUMENTS : BROACHES
Manufacturer (Section D)
DEPUY FRANCE SAS - 3003895575
7 allee irene joliot-curie
b.p. 256
saint priest cedex 69801
FR  69801
MDR Report Key9741990
Report Number1818910-2020-05733
Device Sequence Number18921246
Product Code HTQ
UDI-Device Identifier10603295258186
UDI-Public10603295258186
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 02/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberL20416
Device Catalogue NumberL20416
Device Lot NumberSO2034212
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2020
Initial Date Received by Manufacturer 02/06/2020
Supplement Date Received by Manufacturer03/11/2020
Initial Report FDA Received Date02/21/2020
Supplement Report FDA Received Date03/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age36 YR
-
-