| Model Number |
MMT-723LNAL |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had a compromised force sensor system alarm.The customer did replaced her infusion set but the insulin pump kept going back to fill tubing and unable to continue to the next screen.Customer stated they were unable to clear the alarm and unable to rewound the insulin pump.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test.Device received with motor error alarm during the basic occlusion test due to loose/protruded drive support disk.Unable to perform prime/a33 test, excessive no delivery test and occlusion test or verify a33 alarm due to motor error alarm.
|
| |
|
Search Alerts/Recalls
|