MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC. LIVEWIRE ELECTROPHYSIOLOGY CATHETER DECAPOLAR, MEDIUM SWEEP ELECTRODE SPACING 2; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC. LIVEWIRE ELECTROPHYSIOLOGY CATHETER DECAPOLAR, MEDIUM SWEEP ELECTRODE SPACING 2; CATHETER, STEERABLE Back to Search Results
Model Number 401575
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Cardiac Perforation (2513)
Date of Event 02/04/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The results, method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
Related manufacturer reference number: 2030404-2020-00012, 3008452825-2020-00112, 2182269-2020-00020, 2182269-2020-00021, 3008452825-2020-00113, 2030404-2020-00013, 3005334138-2020-00072.During a pulmonary vein isolation ablation procedure, a pericardial effusion occurred.After the procedure was completed, the patient became hypotensive, so an ultrasound was performed, confirming a pericardial effusion.A pericardiocentesis was performed to stabilize the patient.There were no performance issues with any abbott device.
 
Additional Manufacturer Narrative
One 6f, decapolar, medium sweep, livewire ep catheter was received for evaluation.No visual or function issues were noted.Electrodes 1-10 met specifications of acceptable resistance values with no open or short circuits detected.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported cardiac perforation cannot be conclusively determined.Per the ifu, cardiac perforation is a known risk during the use of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIVEWIRE ELECTROPHYSIOLOGY CATHETER DECAPOLAR, MEDIUM SWEEP ELECTRODE SPACING 2
Common Device Name
CATHETER, STEERABLE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.
14901 deveau place
minnetonka MN 55345
MDR Report Key9744997
Report Number2182269-2020-00022
Device Sequence Number1815637
Product Code DRA
UDI-Device Identifier05414734202183
UDI-Public05414734202183
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K022380
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number401575
Device Catalogue Number401575
Device Lot Number7267714
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/04/2020
Supplement Date Received by Manufacturer03/03/2020
Initial Report FDA Received Date02/24/2020
Supplement Report FDA Received Date03/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
Patient Weight82
-
-