| Model Number |
3286 |
| Medical Device Problem Codes |
Fracture (1260); Material Separation (1562)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
02/04/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event is estimated.Investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 1627487-2020-01871.It was reported that the patient¿s ipg was inoperable and the lead¿s contact plate came off.Additionally, due to patient¿s unrelated health condition the entire system was explanted on (b)(6) 2020 addressing the issue.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of separation was confirmed.As received, the lamitrode was complete.Observed broken and exposed wires at electrode #6, as well as multiple broken wires at the paddle head.The separation and broken wires were consistent with an overstress condition or sudden event the lead was subjected to while still in vivo.(b)(4).The correct information has been updated.
|
| |
|