| Catalog Number |
0035280 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/03/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.
|
| |
|
Event or Problem Description
|
|
It was reported that the black foreign material was found.Per additional information received via email on 17 feb 2020 from ibc representative, it seemed that the foreign material was found inside the syringe.
|
| |
|
Event or Problem Description
|
|
It was reported that the black foreign material was found.Per additional information received via email on 17 feb 2020 from ibc representative, it seemed that the foreign material was found inside the syringe.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was unconfirmed since the product met specifications.Visual evaluation of the returned sample noted one unopened (in original packaging), unused irrigation bulb syringe.It was noted that there was a 0.40 sq mm, brown, cardboard-like speck within the syringe body.This foreign material meets the specification of "loose foreign matter or embedded shall not exceed an aggregate total of 0.6mm² or 1/16¿ per the dirt estimation chart." a device history record review was not required per the investigation.A labeling review was not performed because labelling could not have prevented the reported failure.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|