• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC PROBE 960-556 PLANAR PASSIVE BLUNT; INSTRUMENT, STEREOTAXIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC PROBE 960-556 PLANAR PASSIVE BLUNT; INSTRUMENT, STEREOTAXIC Back to Search Results
Model Number 960-556
Medical Device Problem Codes Mechanical Problem (1384); Material Integrity Problem (2978)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 02/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No products have been returned to medtronic for analysis.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information regarding a navigation system.It was reported that the passive planar had a bent post for one of the spheres.When the sphere was on the post, the instrument would not track due to high geometry error, however, when the sphere was removed and only 4 spheres were on the instrument, it tracked fine verified with no problems.This was reported outside of a procedure during set-up.There was no patient involved in this event.
 
Additional Manufacturer Narrative
H3: the planar passive blunt probe was returned to the manufacturer for analysis.The returned probe displays a high geometry error with markers attached and fully seated.The support arm for marker #5 is bent, not the post, causing the high geometry error.The hardware investigation found that the reported event was related to a hardware issue medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROBE 960-556 PLANAR PASSIVE BLUNT
Common Device Name
INSTRUMENT, STEREOTAXIC
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC
826 coal creek circle
louisville CO 80027
MDR Report Key9746594
Report Number1723170-2020-00620
Device Sequence Number13579181
Product Code HAW
UDI-Device Identifier00673978090122
UDI-Public00673978090122
Combination Product (Y/N)N
Initial Reporter StateCO
Initial Reporter CountryUS
PMA/510(K) Number
K972398
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 03/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number960-556
Device Catalogue Number960-556
Device Lot Number160511
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/20/2020
Supplement Date Received by Manufacturer02/27/2020
Initial Report FDA Received Date02/24/2020
Supplement Report FDA Received Date03/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-