| Model Number |
MICL13.2 |
| Medical Device Problem Codes |
No Apparent Adverse Event (3189); Ejection Problem (4009)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The lens was returned dry in lens vial, wlith residue on lens.Visual inspection found the lens optic and haptic torn.No similar complaint was reported for units within the same lot.(b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated while the surgeon was inserting a 13.2mm micl13.2 implantable collamer lens, - 8.0 diopter, the foam tip plunger overrode the lens and while trying to disengage, the lens was ripped.There was patient contact.The backup lens was implanted.Additional information has been requested but none has been forthcoming.If additional information is received, a supplemental medwatch will be submitted.
|
| |
|