Model Number 20
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Patient Involvement (2645)
Event Date 02/10/2020
Event Type  malfunction  
Manufacturer Narrative
Physio-control performed an initial evaluation of the customer¿s device and verified the reported issue.Physio also observed that the device delivered a monophasic shock and displayed, "abnormal energy delivery".Physio-control continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event Description
The customer contacted physio-control to report that their device logged a critical event code.The event code is indicative of a device issue that could result in a partial loss of defibrillator output energy due to a loss of the negative or positive portion of the biphasic output waveform.The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level.There was no patient use associated with the reported event.
 
Event Description
The customer contacted physio-control to report that their device logged a critical event code.The event code is indicative of a device issue that could result in a partial loss of defibrillator output energy due to a loss of the negative or positive portion of the biphasic output waveform.The defibrillator output energy could be reduced by up to approximately 20% from the selected energy level.There was no patient use associated with the reported event.
 
Manufacturer Narrative
Physio-control sent the customer a quote for repair, however the customer declined to have their device repaired.The customer's device was returned to the customer unrepaired.The cause of the reported issue could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key9748461
MDR Text Key191395200
Report Number0003015876-2020-00260
Device Sequence Number1
Product Code MKJ
UDI-Device Identifier00883873813744
UDI-Public00883873813744
Combination Product (y/n)N
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 04/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number99402-000016
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/14/2020
Date Manufacturer Received04/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-