| Model Number |
144110 |
| Medical Device Problem Codes |
Break (1069); Fracture (1260)
|
| Health Effect - Clinical Codes |
Bone Fracture(s) (1870); Foreign Body In Patient (2687)
|
| Date of Event |
01/27/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As part of normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplement al report will be submitted when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
When trying to remove the bone pins from the patient at the end of the procedure the tibial bone broke.Fluro was called into room to reveal that the bone pin was hit and bent during while performing the tibial keel punch.The surgeon opted to leave the broken pin in the patient.Case type: tka.Surgical delay= 15 minutes.
|
| |
|
Event or Problem Description
|
|
When trying to remove the bone pins from the patient at the end of the procedure the tibial bone broke.Fluro was called into room to reveal that the bone pin was hit and bent during while performing the tibial keel punch.The surgeon opted to leave the broken pin in the patient.Case type: tka surgical delay15 minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: when trying to remove the bone pins from the patient at the end of the procedure the tibial bone broke.Fluro was called into room to reveal that the bone pin was hit and bent during while performing the tibial keel punch.The surgeon opted to leave the broken pin in the patient.Case type: tka, surgical delay<= 15 minutes product evaluation and results: product inspection could not be performed as the product was not returned for evaluation.Product history review: review of the device history records cannot be conducted as the lot number is not provided.Complaint history review: review of the complaint history records cannot be conducted as the lot number is not provided.Conclusions: the failure could not be determined as the product was not available for inspection.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.Corrective action/preventive action: a review of stryker¿s nc/capa database indicated there have been no nc and capa associated with the product and failure mode reported in this event.H3 other text : device not returned.
|
| |
|