• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V200 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V200 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number V200
Device Problem Image Display Error/Artifact (1304)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 25feb2020.
 
Event Description
The customer reported a blurred screen.The service technician confirmed the reported problem, being that the screen could not work normally and the display was abnormal.The service technician restarted the device, checked the service manual, and found the display's abnormal malfunction was caused by the vga (video graphics array) pcb (power control board) failure.The device did not have patient involvement at the time the issue was discovered, therefore, there was no patient or user harm reported.The service technician replaced the vga pcb to resolve the reported issue.The service technician tested and inspected data, and the device turned to full functionally.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V200 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman road
andover MA 01810
Manufacturer Contact
melissa abbott
2271 cosmos court
carlsbad, CA 92011
9093746996
MDR Report Key9751378
MDR Text Key184375862
Report Number2031642-2020-00606
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K102054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/25/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV200
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received02/02/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/18/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-