• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HOLOGIC, INC AFFIRM PRONE BIOPSY SYSTEM; BIOPSY GUIDANCE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HOLOGIC, INC AFFIRM PRONE BIOPSY SYSTEM; BIOPSY GUIDANCE SYSTEM Back to Search Results
Model Number PBX-SYS-AFFIRM-2D
Medical Device Problem Code Unintended System Motion (1430)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 01/31/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As of today the investigation is still in process and a follow up will be filed as needed.
 
Event or Problem Description
It was reported that the table auto move up without pressing any button.No injury reported.A field engineer was dispatched to the site.
 
Event or Problem Description
It was determined that the table control switches needed to be replaced.Once this was completed the system was working as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AFFIRM PRONE BIOPSY SYSTEM
Common Device Name
BIOPSY GUIDANCE SYSTEM
Manufacturer (Section D)
HOLOGIC, INC
36 & 37 apple ridge road
danbury, ct
MDR Report Key9751424
Report Number1220984-2020-00029
Device Sequence Number14160600
Product Code IZH
UDI-Device Identifier15420045506503
UDI-Public(01)15420045506503
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
K153486
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPBX-SYS-AFFIRM-2D
Device Catalogue NumberPBX-SYS-AFFIRM-2D
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/06/2020
Supplement Date Received by Manufacturer03/20/2020
Initial Report FDA Received Date02/25/2020
Supplement Report FDA Received Date03/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-