| Model Number |
TRM-485 |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
01/28/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The reason for reoperation: rupture.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified: underweight, broken shell, wear abrasion, crease flat, red particles.A microscopic analysis was performed which identify a sharp edge opening assessed as unidentified tear opening.A dimension measurement in the shell was performed which identify the thickness within specification, broken striated.Based on the device analysis the final assessment is: a sharp broken on radius due to an unidentified (tear) opening.A striated edge broken on radius side due to surgical damage.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side rupture.Device has been explanted.
|
| |
|