COOK INC GUNTHER TULIP FEMORAL VENA CAVA FILTER SET; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
|
Back to Search Results |
|
Model Number G33016 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problems
Vessel Or Plaque, Device Embedded In (1204); Internal Organ Perforation (1987); Perforation of Vessels (2135)
|
Event Type
Injury
|
Manufacturer Narrative
|
The previous medwatch report was submitted by william cook (b)(4) under manufacturer report reference# 3002808486-2020-00169.Additional information provided determined that this device was manufactured by cook inc.With the submission of this initial medwatch report, cook inc.Informs that all future submissions regarding this complaint will be handled under the manufacturer report reference # referenced in this initial medwatch report.Reporter occupation - non-healthcare professional.(b)(4).Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: vena cava (vc)/organ perforation/embedment, abdominal/chest pain.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Unknown if the reported abdominal/chest pain are directly related to the filter and unable to identify a corresponding failure mode at this point in time.20 devices in lot.No relevant notes on work order.The product is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information become available.
|
|
Event Description
|
Patient allegedly received an implant on (b)(6) 2011 via the right femoral vein due to trauma.Patient is alleging vena cava perforation, organ perforation, and embedment.The patient further alleges "abdominal pain, chest pain on and off but very severe when happens." report from ct: "and inferior vena cava filter is present 4 of the anchoring legs of the filter extend beyond the margin of the wall of the cava.The medial leg extends into the wall of the adjacent aorta.The posterior leg erodes into the anterior inferior margin of the l3 vertebral body in the lateral leg extends slightly into the medial cortex of the right kidney.".
|
|
Event Description
|
27oct2019, report from ct (computed tomography): "ivc filter with the struts extending beyond lumen abutting subjacent loops of bowel, juxtarenal abdominal aorta, and a limited now extending into the l3 vertebral body.".
|
|
Manufacturer Narrative
|
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.The following fields were updated per additional information received: b5, b6, b7, h6.Additional information: investigation investigation is reopened due to additional information provided.Per quality engineering review, the additional information provided for this complaint does not change the previous investigation conclusion.Therefore, no new investigation activities will be conducted at this time.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
|
|
Search Alerts/Recalls
|
|
|