| Model Number |
METER, TRUE METRIX NFRSMG/DL |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Meter was returned for evaluation.No defect was detected.Test strips were not returned for evaluation.Most likely underlying root cause: mlc-58: user had an inaccurate reference: self: the person is using themselves as the reference or how they feel at the time they run the blood test.Note: manufacturer contacted customer in a follow-up call to ensure the replacement products resolved the initial concern - unable to establish contact with customer at this time.
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with test results obtained of 362, 326, 225, 324 and 327 mg/dl.Customer stated that she took insulin shortly after obtaining the result of 362 mg/dl that morning.The customer¿s expected fasting blood glucose test result range is 140 - 200 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.During the call, a blood test was performed by the customer non-fasting and produced test result of 129 mg/dl using meter.The product is stored according to specification in the living room.The test strip lot manufacturer¿s expiration date is 04/30/2021 and test strips have been opened for two months.The meter memory was reviewed for previous test result history: (b)(6).
|
| |
|