| Model Number |
TCF-605 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
04/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iii and patient¿s concern with the product.The device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual and microscopic analysis of the returned device identified: fold creases, curved openings with striated edge on anterior side and a weight test of the explanted material was verified and it was within specification.Based on the device analysis the final assessment is: curved striated openings assessed as surgical damage.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture, baker grade iii and patient¿s concern with the product.The device has been explanted.
|
| |
|