| Model Number |
7060M-100 |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
Undesired Nerve Stimulation (1980)
|
| Date of Event |
02/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Based on the information provided, review of the surgical technique manual, ifu, certificates of analysis and fmea, the most probable root cause is user error; improper implant size selection, using too long of an implant.
|
| |
|
Event or Problem Description
|
|
The patient had right side si joint arthrodesis in (b)(6) 2020 where three implants were installed.The patient complained of right side radicular pain following the procedure.The surgeon determined that the superior positioned implant had breached the neuroforamen.Two days after the initial procedure, the surgeon performed a revision procedure where he removed the implant and replaced it with a shorter implant of the same type to relieve the nerve impingement.No other preexisting implants were adjusted or removed.The patient's radicular pain symptoms resolved following the revision procedure.
|
| |
|
Search Alerts/Recalls
|