• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM MASTERS SERIES VALVE EXPANDED CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 27MECJ-502
Medical Device Problem Codes Migration or Expulsion of Device (1395); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Code Death (1802)
Date of Event 02/02/2020
Type of Reportable Event Death
Additional Manufacturer Narrative
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event or Problem Description
On (b)(6) 2020, a 27mm masters series valve was implanted without any inconveniences and the patient was reported stable post implant.After moving the patient to recovery on (b)(6) 2020, the patient suffered hemodynamic decompensation.An echocardiogram was performed and the physician decided to re-operate.On (b)(6) 2020, during re-operation it was noted that one of the leaflets had dislodged and the other leaflet could not be found.Another 27mm masters series valve was implanted and the patient was moved to recovery.On (b)(6) 2020, the patient had expired and it is unknown if the valve led to the patient's death.Additional information was requested but cannot be obtained.
 
Additional Manufacturer Narrative
An event of leaflet dislodgement following implant was reported.A more comprehensive assessment could not be performed since the device was not returned for analysis, and as attempts abbott medical affairs personnel to obtain more information surrounding the case was unsuccessful.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.This was inclusive of a review of the leaflet and orifice dimensional matching data, which was found to be compliant and conforming with the established dimensional matching requirements.No changes to the spring mechanism for rotation were found to have been made, and the valve was not reprocessed or resterilized.The cause of the leaflet dislodgement could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM MASTERS SERIES VALVE EXPANDED CUFF
Common Device Name
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key9756609
Report Number3003681312-2020-00007
Device Sequence Number2216124
Product Code LWQ
UDI-Device Identifier05414734006453
UDI-Public05414734006453
Combination Product (Y/N)N
Initial Reporter CountryAR
PMA/510(K) Number
P810002
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/08/2023
Device Model Number27MECJ-502
Device Catalogue Number27MECJ-502
Device Lot Number6446628
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 02/04/2020
Supplement Date Received by Manufacturer06/16/2020
Initial Report FDA Received Date02/26/2020
Supplement Report FDA Received Date06/18/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Death; Required Intervention; Disability;
-
-