| Model Number |
TRX-580 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been initiated.If any new, changed or corrected information is noted, a supplemental medwatch will be submitted.The event of "capsular contracture" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture, baker grade unknown.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture, baker grade unknown.Device remains implanted.
|
| |
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.Visual evaluation:visual analysis of the photographs identified: yellow particles on the surface shell.Device analysis performed through photographs, due to the impossibility to perform a further analysis it is not possible to determine the cause of the event.
|
| |
|
Event or Problem Description
|
|
Ccapsular contracture, baker grade iii.Device has been explanted.
|
| |
|
Search Alerts/Recalls
|