MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7/3.5MM VA-LCP ANTEROLATERAL DISTAL TIBIA PL/6 HOLES/LEFT; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.7/3.5MM VA-LCP ANTEROLATERAL DISTAL TIBIA PL/6 HOLES/LEFT; PLATE, FIXATION, BONE Back to Search Results
Model Number 02.118.205
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 02/05/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Initial reporter is synthes sales representative.Without a lot number the device history records review could not be completed.Product was not returned.(b)(4).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, the patient underwent an open reduction internal fixation (orif) of the ankle.During the procedure, the variable angle (va) distal locking screws in the left 6-hole anterior lateral distal tibia plate stripped.Another plate was used and the same thing occurred leaving only three screws in the distal portion of the fracture.There was a one hour surgical delay.The procedure outcome and patient status are unknown.Concomitant device reported: unk - screws: locking (part#: unknown, lot#: unknown, qty 1); 2.7/3.5mm va-lcp anterolateral distal tibia pl/6 holes/left (part#: 02.118.205, lot#: unknown, qty: 1).This report is for one (1) 2.7/3.5mm va-lcp anterolateral distal tibia pl/6 holes/left.This is report 1 of 3 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H3, h4, h6: based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.7/3.5MM VA-LCP ANTEROLATERAL DISTAL TIBIA PL/6 HOLES/LEFT
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9758320
Report Number2939274-2020-01060
Device Sequence Number618612
Product Code HRS
UDI-Device Identifier10886982038364
UDI-Public(01)10886982038364
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K121601
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number02.118.205
Device Catalogue Number02.118.205
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 02/05/2020
Supplement Date Received by Manufacturer02/27/2020
Initial Report FDA Received Date02/26/2020
Supplement Report FDA Received Date03/02/2020
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
2.7/3.5MM VA-LCP ANT DIST TIBIA PL/6 H/L; UNK - SCREWS: LOCKING
Outcome Attributed to Adverse Event Required Intervention;
-
-