• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED PRECICE INTRAMEDULLARY LIMB LENGTHENING SYSTEM; FIXATION DEVICE, INTERNAL, NAIL, INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED PRECICE INTRAMEDULLARY LIMB LENGTHENING SYSTEM; FIXATION DEVICE, INTERNAL, NAIL, INTRAMEDULLARY Back to Search Results
Model Number P10.7-80C245
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 01/31/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No product was returned for evaluation nor were radiographs or images provided to confirm the alleged event.No root cause can be confirmed at this time.
 
Event or Problem Description
Information was received that a revision procedure was performed on (b)(6) 2020.As per reporter the nail was not lengthening.
 
Event or Problem Description
No additional information has been provided.
 
Additional Manufacturer Narrative
Device evaluation: visual inspection revealed no external damage to the returned nail.The x-ray images of the internal components showed no damage and revealed the nail has been distracted beyond the midpoint.The nail was functionally tested and was determined to distract and retract using the high-speed magnet tool as well as an external remote controller (erc).The length as received was measured at 291.38 mm.The maximum and minimum lengths were measured at 328.14 mm and 244.75 mm, respectively, giving a total stroke measurement of 83.39 mm which meets the specification.Distraction force testing was then performed and the nail met the specification.The reported failure could not be confirmed as the nail was found to fully functional.Device records review: a review of the device history records (dhr) indicated that the unit met all required quality specifications and testing prior to product departure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE INTRAMEDULLARY LIMB LENGTHENING SYSTEM
Common Device Name
FIXATION DEVICE, INTERNAL, NAIL, INTRAMEDULLARY
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED
101 enterprise
suite 100
aliso viejo CA 92656
Manufacturer (Section G)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise
suite 100
aliso viejo CA 92656
Manufacturer Contact
tiara rae
101 enterprise
aliso viejo, CA 92656
MDR Report Key9758606
Report Number3006179046-2020-00118
Device Sequence Number688356
Product Code HSB
UDI-Device Identifier00812258020453
UDI-Public812258020453
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K173129
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberP10.7-80C245
Device Lot Number8070208AAA
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/06/2020
Supplement Date Received by Manufacturer04/14/2023
Initial Report FDA Received Date02/26/2020
Supplement Report FDA Received Date04/17/2023
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/02/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ERC
Outcome Attributed to Adverse Event Required Intervention;
Patient Age18 YR
Patient SexMale
Patient Weight48 KG
-
-