• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC SELF-CENTERING BONE FORCEPS 9MM SERRATED JAW-SPEED LOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC SELF-CENTERING BONE FORCEPS 9MM SERRATED JAW-SPEED LOCK Back to Search Results
Model Number 398.79
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Foreign Body In Patient (2687); No Code Available (3191)
Date of Event 01/29/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint #(b)(4).Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.(b)(4).Without a lot number, the device history records review could not be completed as no product was received.The investigation could not be completed; no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, the self-centering bone forcep broke off during surgery.Fragments were generated and was left in the patient.It was unknown if there was a surgical delay.Procedure outcome and patient status are unknown.This complaint involves one (1) device.This report is 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H6: complaint is confirmed as we are able to confirm complaint description (broken) based on the received pictures.H3, h6: the device has been received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h6: investigation summary complaint is confirmed as we are able to confirm complaint description (broken) based on the received pictures.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available (information or/and material), the investigation will be updated as applicable.,it was reported that on (b)(6) 2020, the self-centering bone forcep broke off during surgery.Fragments were generated and was left in the patient.It was unknown if there was a surgical delay.Procedure outcome and patient status are unknown.Flow: damage.Visual inspection: the instrument was received at us cq and upon inspection the portion of the distal jaw with the serrated teeth broke off and was not returned.No other issues were identified.A device failure was identified.Service and repair evaluation: the customer reported the device broke during surgery.The repair technician reported the jaw is broken, pieces are missing.Cutting jaws broken is the reason for repair.The item is not repairable per the inspection sheet.The cause of the issue is unknown.The item will be forwarded to customer quality.The evaluation was confirmed.Document/specification review: the following drawings were reviewed: bone forceps self centering with speed lock: 398_79 rev f.Based on the review of the above drawings, no design issues contributing to relevant complaint condition were identified.Conclusion: no definitive root cause was able to be determined as the circumstances surrounding the complaint are unknown.During the investigation no unidentified product design/manufacturing issues or discrepancies were observed that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SELF-CENTERING BONE FORCEPS 9MM SERRATED JAW-SPEED LOCK
Common Device Name
FORCEPS
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key9758964
Report Number2939274-2020-01064
Device Sequence Number18987108
Product Code HTD
UDI-Device Identifier10886982201638
UDI-Public(01)10886982201638(
Combination Product (Y/N)N
Initial Reporter StateMN
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number398.79
Device Catalogue Number398.79
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/19/2020
Initial Date Received by Manufacturer 02/04/2020
Supplement Date Received by Manufacturer02/19/2020
03/03/2020
Initial Report FDA Received Date02/26/2020
Supplement Report FDA Received Date03/03/2020
03/17/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age22 YR
-
-