| Catalog Number |
3910-500-472 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
02/05/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor remains loose in the patient.
|
| |
|
Event or Problem Description
|
|
It was reported that the anchor remains loose in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.The reported failure mode will be monitored for future reoccurrence alleged failure: anchor remained loose in the patient probable root cause: the reported failure mode will be monitored for future reo"design - inadequate raw material selection for screw - design geometry too sensitive to withstand clinical impaction loads - screw/driver interface not designed to maintain proper assembly process - screw or driver not manufactured to specification - incorrect material used during manufacturing - screw and driver not assembled to specification application - excessive force or leveraging of the screw against the bone - inadequate assessment of bone quality, pilot hole not prepared".
|
| |
|