• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. PLATINUM 1 SERIES; SURGICAL ADJUNCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. PLATINUM 1 SERIES; SURGICAL ADJUNCTS Back to Search Results
Model Number 1MTEC30
Device Problem Failure to Eject (4010)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/18/2019
Event Type  malfunction  
Manufacturer Narrative
Age/date of birth: unknown, information not provided.Gender/sex: unknown, information not provided.Lot number: unknown, information not provided.Unique device identifier (udi #): unknown, as lot number was not provided.Expiration date: unknown, as lot number was not provided.If implanted, give date: not applicable as this is not an implantable device. if explanted, give date: not applicable as this is not an implantable device. device manufacture date: unknown as lot number was not provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that the intraocular lens (iol) had loading issue and was stuck in cartridge, would not advance while inserting.There was patient contact with eye but no medical or surgical intervention were required.Patient is doing fine post-op.No further information provided.
 
Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Manufacturer Narrative
Corrected data: upon further review it was noted in initial mdr submission in section a, a patient date of birth and gender were populated, however this information was actually unknown/not provided.A correction of the information is submitted.Device evaluation: the box of zcb00 lens model was received with the identification of the complaint folder.However, only the lens was in the box.There was no cartridge received.Therefore, the reported issues could not be verified, and product quality deficiency could not be determined.Manufacturing records review: the manufacturing record evaluation could not be performed because the lot number of the complaint product is unknown and is not available for evaluation.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson and johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM 1 SERIES
Type of Device
SURGICAL ADJUNCTS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key9759622
MDR Text Key181345506
Report Number2648035-2020-00183
Device Sequence Number1
Product Code KYB
UDI-Device Identifier05050474540323
UDI-Public(01)05050474540323(10)UNKNOWN
Combination Product (y/n)Y
PMA/PMN Number
K081545
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup,Followup
Report Date 01/01/2005,11/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1MTEC30
Device Catalogue Number1MTEC30
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Date Manufacturer Received10/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ZCB00 LENS, SN: (B)(6)
-
-