| Model Number |
GIA8038RL |
| Medical Device Problem Codes |
Loss of or Failure to Bond (1068); Failure to Form Staple (2579)
|
| Health Effect - Clinical Codes |
Death (1802); Tissue Damage (2104); No Code Available (3191)
|
| Date of Event |
01/21/2020
|
|
Type of Reportable Event
|
Death
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, post-operatively, on an open hemicolectomy, the staples had failed to close the staple line at the colon because of poor staple formation.The patient went back for reoperation to resolve the issue in order to complete the case.Hence, there was tissue damage.The patient died.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Additional Manufacturer Narrative
|
|
H3 evaluation summary: post market vigilance (pmv) led an evaluation of one photograph and one representative device.Photographic inspection noted tissue and the staples appeared fully formed.In the photograph it appeared the tissue on one side of the staple line was damaged and ripped.Visual inspection of the returned instrument noted no abnormalities.The loading unit was received sealed in the blister package with no abnormalities noted.The loading unit was removed from the blister package for functional testing.The loading unit was then loaded into the returned instrument.The instrument and loading unit was applied to the appropriate test media with acceptable results.Proper staple formation was observed and the knife blade cut both completely and cleanly.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.The investigation concluded there were no abnormalities that would have caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|