| Model Number |
MMT-751LNAL |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Difficult to Program or Calibrate (1496); Obstruction of Flow (2423); Battery Problem (2885); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
02/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the buttons were harder to press.Customer stated that the insulin pump had no delivery alarms.Customer stated that the insulin pump had unusual noises different from the normal rewind noise.Rewinding was slower.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with intermittent button response due to partial unlocked lcd keypad connector noted during visually inspect.Unable to perform all functional testing including the displacement, operating currents, a21 error test, rewind, basic occlusion, occlusion, prime or a33 and excessive no delivery alarm or verify failed battery test alarm, charge or battery lasts less than excepted anomaly and rewind anomaly due to buttons not responding.(b)(4).
|
| |
|