| Model Number |
FF-410335 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Pain (1994)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-001-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: analysis of the returned device identified: opening on posterior and underweight.A visual and microscopic analysis was performed which identified: sharp opening and crease fold.A dimension measurement in the shell was performed which identify the thickness within specification.Based on the device analysis the final assessment is: a sharp opening on posterior due to an unidentified (tear) opening the event of "pain" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: pain and patient's concern for the product.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side "exchange from textured to smooth breast implants due to the patient¿s concern with the product" and "pain".The device was explanted.
|
| |
|