• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS LEGEND DISSECTING TOOL; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS LEGEND DISSECTING TOOL; DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) Back to Search Results
Model Number 8TA11
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 12/20/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No conclusion can be drawn.No evaluation was performed, as the device was not returned.If the device is returned in the future, product analysis may be performed.The user manual contains the following warning ¿do not use excessive pressure, such as bending or prying, on attachments or dissecting tools.This may cause tool to bend or break and cause injury to patient, operator and/or operating room staff.¿ we will continue to track and trend this complaint type.Initial reporter phone number: (b)(6).Medwatch report number: mw5092708.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the tip of the tool was broken after use.It was also added that the broken piece was not found in either the surgical field or surgical site.No patient impact reported.
 
Additional Manufacturer Narrative
Phone number: (b)(6).If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND DISSECTING TOOL
Common Device Name
DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED)
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
MDR Report Key9765518
Report Number1625507-2020-00026
Device Sequence Number572514
Product Code HBE
UDI-Device Identifier00643169765689
UDI-Public00643169765689
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/06/2024
Device Model Number8TA11
Device Catalogue Number8TA11
Device Lot Number0218616053
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/30/2020
Supplement Date Received by Manufacturer03/03/2020
Initial Report FDA Received Date02/27/2020
Supplement Report FDA Received Date03/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-