| Model Number |
A2025-120 |
| Medical Device Problem Code |
Device Markings/Labelling Problem (2911)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
02/06/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device has been received.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was reported that the box for an armada 14 was pulled and just before use, the device was noted to be an armada 35 percutaneous transluminal angioplasty (pta) catheter.The discrepancy was noted before use and the device was not used.A new device was pulled and was able to be successfully used.Reportedly, the account may have inadvertently placed the wrong device in the box but that could not be confirmed.There was no clinically significant delay in the procedure.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for analysis.The reported product discrepancy was confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents reported from this lot.There is no indication of a product quality issue with respect to manufacture, design or labeling.The investigation determined the reported/noted difficulties appear to be related to circumstances of the procedure as it is likely that a mix-up occurred at the hospital.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|