• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. PLATINUM 1 SERIES; SURGICAL ADJUNCTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. PLATINUM 1 SERIES; SURGICAL ADJUNCTS Back to Search Results
Model Number 1MTEC30
Device Problem Failure to Eject (4010)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/07/2019
Event Type  malfunction  
Manufacturer Narrative
Lot number: unknown, as the lot number was not provided.Expiration date: unknown as product lot number was not provided.If implanted, give date: not applicable, as the cartridge is not an implantable device.If explanted, give date: not applicable, as the cartridge is not an implantable device.Device manufacture date: unknown, as the lot number was not provided.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event Description
It was reported that an intraocular lens (iol) was stuck in the cartridge, and the cartridge broke-cracked, the lens was partially inserted into the patient's right eye.The same model and diopter lens was implanted as a replacement.The patient has recovered.No additional information was provided to johnson & johnson surgical vision.
 
Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Manufacturer Narrative
Device evaluation: the device was not returned at the manufacturing site; therefore; product testing could not be performed, and the customer¿s reported complaint could not be verified.Manufacturing record review: the manufacturing record could not be reviewed since the lot number was not provided.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINUM 1 SERIES
Type of Device
SURGICAL ADJUNCTS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key9767787
MDR Text Key181536593
Report Number2648035-2020-00191
Device Sequence Number1
Product Code KYB
UDI-Device Identifier05050474540323
UDI-Public(01)05050474540323(10)UNKNOWN
Combination Product (y/n)Y
PMA/PMN Number
K081545
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup,Followup
Report Date 01/01/2005,11/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number1MTEC30
Device Catalogue Number1MTEC30
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Report Sent to FDA01/01/2005
Date Report to Manufacturer01/10/2005
Initial Date Manufacturer Received 01/31/2020
Initial Date FDA Received02/28/2020
Supplement Dates Manufacturer Received04/17/2020
10/25/2020
Supplement Dates FDA Received05/12/2020
11/02/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
MODEL ZCB00 LENS SN: (B)(6)
Patient Age67 YR
-
-