|
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 3007963827-2020-00060, 3007963827-2020-00061.Concomitant medical products: femur cemented cruciate retaining (cr) narrow right, item#: 42502005402, lot#: 63405040.Tibia cemented 5 degree stemmed right, item#: 42532006402, lot#: 63846697.Report source - (b)(6).Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
This follow-up is being submitted to relay additional information.Updated: b4, b5, g4, g7, h1, h2, h3, h6, and h10.No product was returned or pictures provided; therefore, visual and dimensional evaluations could not be performed.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: patella baja is defined by the insall-salvati ratio (patellar tendon length-length of the posterior surface of the tendon from the lower pole of the patella to its insertion on the tibial tubercle) divided by (patellar length-greatest pole-topole length).In this provided image, the ratio measures 0.817.Patella baja occurs when the ratio measures less than 0.8, indicating borderline patella baja.Overall fit and alignment of the implants, as well as bone quality, appear normal.One-page clinic note is for initial surgical planning and scheduling.No other information was provided.Patella baja was confirmed however, pain and stiffness issues cannot be confirmed.Device history record (dhr) was reviewed for deviations and/ or anomalies with no related deviations / anomalies identified.A definitive root cause cannot be determined.Per package insert 87-6204-022-23 persona the personalized knee system: pain is known adverse effect of the system.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|