Model Number 3772
Device Problems Disconnection (1171); High impedance (1291)
Patient Problem Inadequate Pain Relief (2388)
Event Date 02/10/2020
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
Manufacturer report number 1627487-2020-02141.It was reported that post ipg replacement procedure (reference reg report: 1627487-2020-01227), high impedances were diagnosed on multiple lead contacts.X-ray confirmed that the lead tail was disconnected from the new ipg.As a result, surgical intervention took place on (b)(6) 2020 wherein the ipg was explanted, replaced and connected to the existing lead.Effective therapy was established post op.
 
Manufacturer Narrative
The reported event for high impedances was not confirmed.Visual inspection of the returned ipg did not reveal any anomalies that would contribute to the complaint.The ipg passed a lead fitment test as well as lead retention test for both header chambers.The ipg was functionally tested and passed all testing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRODIGY MRI IPG, 16CH, RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION (PUERTO RICO, LLC)
lot a interior - #2 street km 67.5
santana industrial park
arecibo PR 00612
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key9769754
MDR Text Key181469834
Report Number3006705815-2020-00916
Device Sequence Number1
Product Code LGW
UDI-Device Identifier05415067017093
UDI-Public05415067017093
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/12/2020
Device Model Number3772
Device Catalogue Number3772
Device Lot NumberA000070642
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/25/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/10/2020
Initial Date FDA Received02/28/2020
Supplement Dates Manufacturer Received03/24/2020
Supplement Dates FDA Received03/30/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/13/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
SCS LEAD
Patient Outcome(s) Other;
Patient Weight68
-
-