Model Number ZA9003
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/28/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Device evaluation: the product returned was received at site.The haptic detached was observed.Viscoelastic residues and particles were in lens surface.The condition is consistent with the implanted/explanted process.Due the conditions of the returned sample product deficiency could not be determined.Manufacturing record review: the manufacturing process record was evaluated and revealed that the product was manufactured and released according to specifications.A search revealed that no additional investigation request were received from this production order.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a za9003 intraocular lens (iol) when being inserted into the eye was removed.It was learned however, upon further follow-up that the iol was found to have dislocated during the post-op exam.It was confirmed that how the event occurred was that the patient was implanted the lens, went out of operating room (or) but had to go back into the or during the same day, because the doctor said that the lens dislocated.Patient went back into the or for lens removal.Patient post-op was uneventful, no injuries, no adverse events.No other information was available.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key9770702
Report Number2648035-2020-00195
Device Sequence Number16874224
Product Code HQL
UDI-Device Identifier05050474528970
UDI-Public(01)05050474528970(17)231024
Combination Product (Y/N)Y
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/04/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/24/2023
Device Model NumberZA9003
Device Catalogue NumberZA90030180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/21/2020
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 02/13/2020
Supplement Date Received by Manufacturer10/25/2020
Initial Report FDA Received Date02/28/2020
Supplement Report FDA Received Date11/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age38 YR
Date Report Sent to FDA01/01/2005
-
-