• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOPERSURGICAL, INC. LEEP SYSTEM 1000 ESU GEN.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOPERSURGICAL, INC. LEEP SYSTEM 1000 ESU GEN. Back to Search Results
Model Number 52969
Device Problem Output below Specifications (3004)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/18/2020
Event Type  malfunction  
Manufacturer Narrative
Coopersurgical inc.Is currently investigating the reported condition.
 
Event Description
Customer stated "low output".Reference repair order #: (b)(4).
 
Manufacturer Narrative
Investigation: x-inspect returned samples.Analysis and findings: complaint #(b)(4).Distribution history: this complaint unit was manufactured at csi on 8/17/2005 under wo #(b)(4) and shipped on 08/24/2005.Manufacturing record review: a review of the device history record could not be performed as the record could not be located at this time.However, at the time of manufacture, records from each unit are reviewed to ensure that product meet all specifications.Should the device history record be located, this complaint will be amended accordingly.Incoming inspection review: n/a.Service history record: no additional service history records found for this unit.Historical complaint review: a review of the 2-year complaint history showed similar reported complaint conditions.Product receipt: the complaint unit was returned on a repair.However, based on log 93909, this unit was at csi on 2/19/2020.Visual evaluation: visual examination of the complaint unit revealed no physical damage.Functional evaluation: complaint unit was functionally evaluated and found to function properly.Root cause: the product tested to specification as the device was found to meet all visual and functional test specifications.Updated component: the diaphragm was changed out due to a previous finding where the material degraded to the point where it no longer functioned as intended and prevents activation of a unit.This unit's diaphragm was in working order.However, the corrective action for this issue requires returned units be updated to the new material and applicable on this unit.Correction and/or corrective action: coopersurgical will continue to monitor this complaint condition for trends.No further corrective action is necessary, as the complaint condition was not confirmed.Corrective actions relevant to the updated diaphragm are as follows; coopersurgical service and repair replaced the diaphragm on the unit, tested it and returned it to the customer.Engineering has successfully tested a replacement material made of silicone for use in repairs going forward, eng-test-10341-r.The ifu was also updated to add a safety check via ecn-20444.A service bulletin was issued to existing customers informing them to check for this issue and return the unit if needed.All product in fg and sk, as applicable, were reworked to replace the previous versions of the dfus on all applicable products.The repaired unit will have been repaired with this new silicone material.Preventative action activity : coopersurgical will continue to monitor this complaint condition for trends.
 
Event Description
Customer stated "low output" reference repair order #(b)(4).(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEEP SYSTEM 1000 ESU GEN.
Type of Device
LEEP SYSTEM 1000 ESU GEN.
Manufacturer (Section D)
COOPERSURGICAL, INC.
95 corporate drive
trumbull, ct
MDR Report Key9771125
MDR Text Key191403578
Report Number1216677-2020-00069
Device Sequence Number1
Product Code HGI
Combination Product (y/n)N
PMA/PMN Number
K952483DISCO
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Remedial Action Repair
Type of Report Initial,Followup
Report Date 05/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number52969
Device Catalogue Number52969
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/24/2020
Initial Date FDA Received02/28/2020
Supplement Dates Manufacturer Received02/24/2020
Supplement Dates FDA Received06/01/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-