| Model Number |
R SERIES |
| Medical Device Problem Codes |
Pacing Problem (1439); Defibrillation/Stimulation Problem (1573)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during a routine shift check by a clinician, the device displayed "defib fault 72" and "pacer fault 115" messages.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation.Instead, the suspect pace/defib engine board was returned.The customer's report was observed and attributed to a faulty integrated circuit on the pace/defib engine board.The board was scrapped.Analysis of reports of this type has not identified an increase in trend.
|
| |
|
Search Alerts/Recalls
|